The model that was right in the laboratory and wrong in the waiting room

The artificial intelligence tools now reaching the eye examination, what each one does, and how little of the trial evidence measures vision.

In this issue6 sections · about 16 minutes

Between August 2021 and June 2022, researchers from the Centre for Eye Research Australia placed a self-imaging fundus camera in two general practice clinics in Victoria. An artificial intelligence (AI) model read each photograph for glaucoma. General practices do not normally own a fundus camera, and the authors chose the setting for that reason. They write that general practitioners "typically lack both the equipment and training necessary to diagnose glaucoma" [1]. Patients were approached while they waited to see their doctor.

The model was first trained in 2016 on colour fundus photographs from a Chinese population. In its original validation it reached a sensitivity of 95.6 percent and a specificity of 92.0 percent. However, it did less well in the two clinics.

  • One in three people could not be photographed. Of 414 adults aged 50 or over who consented, 277 had an image captured from at least one eye. The people who could not be imaged were older, 71.4 years against 64.8, and more often reported a cataract diagnosis, 40.9 percent against 19.1 percent.
  • Sensitivity was 65.0 percent and specificity was 94.6 percent, counted per person. The reference standard was two therapeutically endorsed optometrists grading the same photographs, with an ophthalmologist adjudicating. It included no optical coherence tomography (OCT) and no visual field test.
  • Most flagged patients did not attend. Among 161 people who had seen an eye care provider in the previous twelve months with no glaucoma diagnosis, the model flagged 18. Sixteen could be contacted. Six attended. One reported a glaucoma diagnosis at telephone follow-up.

The authors wrote that "too many patients in our study did not adhere to referral" [1].

Figure 1. The same glaucoma model, in the laboratory and in two Victorian general practicesSensitivity was 95.6 percent in the original validation and 65.0 percent in the clinics. Of 161 people with recent eye care and no glaucoma diagnosis, 18 were flagged, 6 attended and 1 reported a diagnosis.
A. Performance, percent
Original validationClinic, with exact 95 percent interval from the published counts
Sensitivity
Specificity
0255075100
B. Everyone who consented, people
Consented
414
Image captured, at least one eye
277
C. Subgroup: seen by an eye care provider in the previous 12 months, no glaucoma diagnosis, people
People in this subgroup
161
Flagged as referable by the AI
18
Contactable
16
Attended the referral
6
Reported a glaucoma diagnosis at follow-up
1
Jan CL et al, npj Digital Medicine 2025. Validation figures are the model's published performance in its original validation, in a different population, so the gap is not the effect of deployment alone. Clinic figures are per person, across the 277 people with an image captured from at least one eye, against two therapeutically endorsed optometrists grading the same photographs with ophthalmologist adjudication. The clinic intervals are exact binomial intervals I calculated from the paper's cross-tabulation, 13 of 20 for sensitivity and 243 of 257 for specificity. Those counts reproduce the paper's sensitivity, specificity and predictive values. They count 5 people who were ungradable on manual grading as referable, and the methods say such participants were excluded. The paper prints 59.4 to 70.6 and 91.9 to 97.2. Panel C is a subgroup and not the next stage for everyone in panel B: the 161 people who had seen an eye care provider in the previous twelve months with no glaucoma diagnosis. The referral chain is reported for that group only, and the one diagnosis is a patient report at telephone follow-up. Bars shorter than 0.6 percent of the scale are drawn at 0.6 percent so that they stay visible. [1]
Data as a table
GroupStagePeople
Everyone who consentedConsented414
Everyone who consentedImage captured, at least one eye277
Subgroup: eye care in 12 months, no glaucoma diagnosisPeople in this subgroup161
Subgroup: eye care in 12 months, no glaucoma diagnosisFlagged as referable by the AI18
Subgroup: eye care in 12 months, no glaucoma diagnosisContactable16
Subgroup: eye care in 12 months, no glaucoma diagnosisAttended the referral6
Subgroup: eye care in 12 months, no glaucoma diagnosisReported a glaucoma diagnosis at follow-up1
MeasureOriginal validationClinicExact 95% interval, clinicInterval as printed
Sensitivity95.665.040.8 to 84.659.4 to 70.6
Specificity92.094.691.0 to 97.091.9 to 97.2

The study appeared in npj Digital Medicine in 2025. The title calls it a pragmatic trial. The authors describe the design as prospective observational, with no comparator arm and no trial registration. They declare that one senior author is chief medical officer of Eyetelligence, an Australian retinal AI company. It is still the paper I would hand a colleague first, because it reports what product literature leaves out. It counts who could not be imaged, what the sensitivity was in a real clinic, and how many flagged patients turned up.

In my own practice, the referrals I remember are the ones that came back marked "did not attend".

AI is sold to optometrists on its accuracy, and this series looks at what happens after that number. A tool can be accurate in a validation study and still not improve care in a primary care practice. A tool can also be in wide use without any evidence that it helps patients. Both apply to the tool I sell, and I come back to that at the end.

One number in the paper does not reproduce

The sensitivity of 65.0 percent is printed with a 95 percent confidence interval (CI) of 59.4 to 70.6, and the specificity with 91.9 to 97.2. The paper's own table shows 15 people with referable glaucoma on manual grading and 5 more whose photographs the graders could not grade. The published figures reproduce from that table only when those 5 are counted as referable, which makes the sensitivity 13 of 20. The methods say that such participants were excluded, so the table and the methods do not agree.

An exact interval on 13 of 20 runs from 40.8 to 84.6 percent by my calculation. The published interval has the width that a denominator of 277 would give. Only the authors can settle it. Until they do, the true sensitivity in those clinics could plausibly be anywhere from about 40 to 85 percent.

Where AI now sits in an eye examination

The glaucoma camera is one tool, used at one point in an eye examination. Most of the AI now sold to optometrists and ophthalmologists comes in at one of five points.

  • Before the clinician: software on a fundus camera reads the photograph and returns a referral decision during the visit, with no clinician looking at the image.
  • Beside the clinician: software reads the photograph or OCT scan the clinician has already taken, and adds a diagnosis, a referral suggestion or a measurement.
  • Past the eye: software reads the retina for cardiovascular risk, or for the gene behind an inherited retinal disease.
  • At the phoropter and in surgical planning: a deep learning refraction, and an intraocular lens (IOL) calculator.
  • With the patient: a chatbot that answers questions about an operation.

Before the patient reaches a clinician

The tool is software attached to a fundus camera, usually in a general practice, diabetes or endocrinology clinic. It reads the photograph for diabetic retinopathy and returns a result while the patient is still there. The person taking the photograph does not have to be an eye care professional. In the Pilbara, Lions Outback Vision ran Google's Automated Retinal Disease Assessment (ARDA) from a van. A research officer with no formal healthcare qualification took the photographs, and patients were told the result on the spot [2].

Four systems are cleared by the United States (US) Food and Drug Administration (FDA) to make the decision without a clinician. They are LumineticsCore from Digital Diagnostics, formerly IDx-DR, EyeArt from Eyenuk, AEYE-DS from AEYE Health, and iPredict-DR from iHealthscreen, which was cleared on 2 July 2026 [3, 4].

The examination moves to where the patient already is, and the result comes back during that visit instead of at an eye appointment.

  • In a US trial, 100 percent of young people with diabetes offered the AI examination in their own diabetes clinic completed an eye examination within six months. Of those given a scripted referral to an eye care provider and education, 22 percent did [5].
  • In specialist eye clinics in Bangladesh, clinic days randomised to an FDA-authorised autonomous system completed 1.59 patient encounters per specialist hour, against 1.14 without it [6].
  • In England, eight CE-marked systems were tested on the stored images from 202,886 encounters in the North East London screening programme, against the programme's own human grading. The question was whether software could take the first read and leave the human graders only the images it flags. Sensitivity for moderate to severe retinopathy ran from 96.7 to 99.8 percent across the eight. The false-positive rate ran from 4.3 to 61.4 percent, so the choice of vendor would decide how much human grading the filter saves [7].

For an optometrist, the referral in this pathway arrives from a diabetes or general practice clinic, and the finding behind it is the software's. The patient still has to make the trip: in the US trial, 16 of 25 young people with an abnormal result followed through to an eye care provider [5]. The next issue follows these systems out of their pivotal trials.

Beside the clinician

The tool reads the photograph or scan the clinician has already taken and shows its own reading beside the clinician's. The clinician still signs the decision.

  • EyeFM is a vision-language model pretrained on 14.5 million ocular images from five imaging modalities, paired with clinical text. In a single-centre, double-masked randomised trial in a Chinese high-risk screening population, 668 participants were allocated to 16 ophthalmologists, half working with the copilot and half without. Ophthalmologists with it reached the correct diagnosis in 92.2 percent of cases against 75.4 percent, and the correct referral in 92.2 against 80.5 percent. Their reports also scored higher on a standardisation measure. At follow-up, 33.7 percent of patients in the copilot arm had acted on a referral suggestion, against 20.2 percent [8].
  • The Hong Kong AI-OCT system adds a macular oedema reading to the fundus photograph report a screening programme already produces. It checks image quality, detects diabetic macular oedema and flags the scans it is unsure of. The trial was multicentre, randomised and designed for non-inferiority, in 276 patients referred from the territory-wide screening programme. Referring on both reports cut false-positive referrals to 24.1 percent, against 69.1 percent on the photograph report alone. Sensitivity was 100 percent in both arms [9]. For ethical reasons every participant still saw the specialist, so the trial compared referral decisions and nothing downstream of them.
  • Two RetInSight products on the Australian register take measurements instead of making a diagnosis. The Fluid Monitor quantifies the volume of intraretinal fluid, subretinal fluid and pigment epithelial detachment on an OCT scan, to support treatment decisions in neovascular age-related macular degeneration. The GA Monitor, GA standing for geographic atrophy, quantifies the area of retinal pigment epithelium loss and photoreceptor degeneration in geographic atrophy. The register entry says the results must not be the sole basis for a diagnosis or a treatment decision [10]. I found no trial of either in the search window.
  • The one copilot tested in community optometry did not do well. In the United Kingdom HERMES study, an OCT referral model called Octane was run on scans from community optometrists referring retinal disease. It could process images from 204 of 396 participants, 51.5 percent. On referral decisions it was less accurate than hospital experts and of similar accuracy to the community optometrists [11].

In the two trials the second reading raised the share of right diagnoses and cut the share of needless referrals. Two other studies looked at what it does to the clinician's own judgement. I come back to them after the gaps.

Beyond the eye

  • Mediwhale's software on the Australian register takes a fundus image and indicates suspected cataract, glaucoma and retinal disease, and predicts cardiovascular risk. Both Eyetelligence entries on the register also carry a cardiovascular risk output [10]. I found no randomised trial of cardiovascular risk read from the retina in the search window, so what it changes for a patient is not yet known.
  • Retina4IRD reads colour fundus photographs and OCT scans and predicts which of 17 genotype categories lies behind a suspected inherited retinal disease, before genetic testing. In a multicentre trial, 300 people with suspected disease were randomised to a specialist with or without it, and 295 had sequencing reports to compare against. The correct genetic category was in the top five predictions in 88.5 percent of cases with the tool, against 67.3 percent without. The top prediction was right in 37.8 against 22.4 percent [12]. In practice the specialist would order the genetic test with a shortlist already in hand.

Refraction and surgical planning

  • Tongren Digital Sight is a portable autorefractor: an infrared photorefraction camera and a deep learning model trained on 362,000 images. In a randomised crossover trial of 282 adults at Beijing Tongren Hospital, its spherical equivalent differed from subjective refraction by a mean of 0.07 dioptres. Corrected acuity with its prescription was non-inferior on the day, with a mean difference of 0.03 on the logMAR (logarithm of the minimum angle of resolution) scale. Agreement was weaker in high myopia [13]. The authors built it for places with no phoropter and no one trained to use one. Every participant in the trial also had a subjective refraction to compare it with.
  • The Zeiss calculator on the Australian register predicts the post-operative residual refraction for a given eye biometry and IOL power. It also helps the surgeon choose the lens model and its axis for corneal astigmatism [10]. It is listed as AI-enabled. No trial of it appears in the search window.

With the patient

  • A cataract surgery chatbot at the university hospital in Brussels answered only with clinician-validated responses. In a pilot randomised trial of 64 patients, brochures plus the chatbot made no detectable difference to knowledge, anxiety or satisfaction compared with brochures alone. Only 35 of the 64 returned the questionnaire. Seventeen of the 33 patients given the chatbot never used it, and those who did were younger on average, 64 years against 74 [14]. A pilot this small cannot show that the two approaches are equivalent either.

What none of this has shown

No trial followed vision over time. I found seven randomised trials of an AI tool in ophthalmology or optometry published between January 2023 and September 2026 [5, 6, 9, 12, 8, 13, 14]. None had a vision outcome as its primary endpoint. Three measured a process of care, three measured accuracy or agreement, and one measured what patients knew and how anxious they were. The autorefraction trial measured corrected acuity on the day as a secondary outcome, and it was non-inferior.

Figure 2. The primary endpoint of each of the seven randomised trials I found, January 2023 to September 2026Three process endpoints, three accuracy or agreement endpoints, one patient-reported endpoint. No trial had vision as its primary endpoint. One measured same-day corrected acuity as a secondary outcome.
Process of care3 trials
ACCESS164 young people. Two paediatric diabetes sites, Baltimore.
Eye examination completed within 6 months100 vs 22 percent
B-PRODUCTIVE993 patients, 105 clinic days. Retina clinics, Bangladesh, randomised by clinic day.
Completed encounters per specialist hour1.59 vs 1.14
AI-OCT for macular oedema276 patients. Multicentre, Hong Kong screening programme.
False-positive referral rate24.1 vs 69.1 percent
Accuracy or agreement3 trials
Retina4IRD300 randomised, 295 analysed. Multicentre, inherited retinal disease.
Top-5 genetic category accuracy88.5 vs 67.3 percent
EyeFM copilot668 participants. Single centre, 16 ophthalmologists, China.
Correct diagnosis, and correct referral92.2 vs 75.4 percent, and 92.2 vs 80.5 percent
Deep-learning autorefraction282 adults. Crossover, Beijing Tongren Hospital.
Agreement with subjective refraction, spherical equivalentMean difference 0.07 dioptres, r = 0.96
Patient-reported1 trial
Cataract education chatbot64 patients. Single-centre pilot, Brussels.
Knowledge, anxiety and satisfactionNo significant difference detected
Vision kept over time0 trials
No trial. The autorefraction trial measured corrected acuity on the day as a secondary outcome, and it was non-inferior to subjective refraction.
Trials that randomised patients or clinic days to an AI-containing pathway and were published in the window, found by the searches described in the text. HERMES is excluded because it randomised teleophthalmology, not the AI. Each result is the AI arm against the comparator on the trial's stated primary endpoint, except the autorefraction trial, where it is the mean difference from subjective refraction and the correlation with it. Secondary outcomes were not charted for every trial, so the empty group means that no trial had vision as its primary endpoint, not that vision was never measured. The grouping by endpoint type is mine. The chatbot pilot detected no significant difference in 64 patients, which is not evidence that the two arms were equivalent. [5, 6, 8, 9, 12, 13, 14]
Data as a table
TrialParticipantsSettingPrimary endpointResult, AI arm vs comparatorEndpoint group
ACCESS164 young peopleTwo paediatric diabetes sites, BaltimoreEye examination completed within 6 months100 vs 22 percentProcess of care
B-PRODUCTIVE993 patients, 105 clinic daysRetina clinics, Bangladesh, randomised by clinic dayCompleted encounters per specialist hour1.59 vs 1.14Process of care
AI-OCT for macular oedema276 patientsMulticentre, Hong Kong screening programmeFalse-positive referral rate24.1 vs 69.1 percentProcess of care
Retina4IRD300 randomised, 295 analysedMulticentre, inherited retinal diseaseTop-5 genetic category accuracy88.5 vs 67.3 percentAccuracy or agreement
EyeFM copilot668 participantsSingle centre, 16 ophthalmologists, ChinaCorrect diagnosis, and correct referral92.2 vs 75.4 percent, and 92.2 vs 80.5 percentAccuracy or agreement
Deep-learning autorefraction282 adultsCrossover, Beijing Tongren HospitalAgreement with subjective refraction, spherical equivalentMean difference 0.07 dioptres, r = 0.96Accuracy or agreement
Cataract education chatbot64 patientsSingle-centre pilot, BrusselsKnowledge, anxiety and satisfactionNo significant difference detectedPatient-reported

The wider AI trial literature to early 2024 looks much the same. A scoping review of 86 randomised trials of AI in clinical practice found that 70, or 81 percent, had positive primary endpoints. In 46 of the 86, or 54 percent, the endpoint was diagnostic yield or performance. Its full text records that all three ophthalmology trials reported a reduction in operational time [15].

The count comes from the evidence map behind this series. The map was built from PubMed, ClinicalTrials.gov, the Australian New Zealand Clinical Trials Registry and the regulator lists between 27 and 30 September 2026. It is not a registered systematic search, so a trial indexed under terms I did not use would be missing.

Two more gaps run through the sections above. Attendance after a positive result is the step the software cannot take for the patient: 16 of 25 in the US trial, 6 of 18 in the glaucoma clinics, and one in three in the EyeFM copilot arm. And outside diabetic retinopathy the evidence is thin. The glaucoma model that opened this issue was far less sensitive in general practice than in its original validation. The one OCT copilot tested in community optometry was no more accurate than the optometrists.

When the clinician and the AI disagree

Two studies measured what an AI reading does to the clinician's own decision.

Explaining a glaucoma AI's recommendation did not measurably improve optometrists' accuracy, and it increased reliance on the AI's mistakes. One user study from the search period put 87 optometrists in front of a glaucoma referral model built from routine eye care data. Accuracy was 51 percent unaided (95 percent CI 46 to 56). It was 58 percent with the AI recommendation alone, 60 percent with a feature-importance explanation and 62 percent with a scorecard explanation. The three AI-supported conditions did not differ significantly from one another, so the study found no gain from the explanations. Post-hoc explanations "increased over-reliance on incorrect AI recommendations" [16].

The AI was right on 80 percent of the study cases. Those were 20 cases selected for the study, so the 80 percent describes the case set and not the model in a clinic.

A second study asked 45 trainee and expert ophthalmologists to grade 120 fundus photographs. A week later they read the same photographs again with an AI whose judgements were deliberately wrong 30 percent of the time. Accuracy rose from 48.4 to 59.6 percent overall. It was 63.9 percent when the AI was right and 47.9 percent when it was wrong [17].

However, the paper tested the readings made with a wrong AI against the first-session readings and found no significant difference. The 47.9 percent covers a subset of the photographs and the 48.4 percent covers all 120, so those two figures are not a matched pair. The gain was larger for trainees, 14.2 percentage points against 8.6, and the effect of a wrong AI was less pronounced in experts. The standard deviations run from 20 to 27 percentage points, and a repeat reading a week apart carries its own learning effect.

Figure 3. Clinicians with AI: better on average, with no significant gain when the AI was wrongOptometrists with AI scored below the AI on the study cases. Ophthalmologists scored 63.9 percent when the AI was right and 47.9 percent when it was wrong.
A. 87 optometrists, urgent glaucoma referral, percent correct
B. 45 ophthalmologists, glaucoma grading, all 120 photographs, percent correct
C. The second reading, split by whether the AI was right, percent correct
Panel A: a user study of 87 optometrists with a glaucoma referral model. The published 95 percent intervals are 46 to 56 unaided, 56 to 60 with no explanation, 55 to 64 with feature importance and 57 to 66 with a scorecard, and the three AI-supported conditions did not differ significantly. The AI figure is its accuracy on the 20 cases selected for the study (Gomez 2025). Panels B and C: 45 trainee and expert ophthalmologists grading 120 fundus photographs unaided, then the same photographs one week later with an AI whose judgements were deliberately wrong 30 percent of the time. Standard deviations are 24.8, 20.3, 20.6 and 26.6 points in the order shown, and a standard deviation is not a confidence interval (Kashiwagi 2025). Panel C splits the second reading into two subsets. The 47.9 percent is on the photographs where the AI was wrong and the 48.4 percent is on all 120, so the two are not a matched pair. The paper's own test of the wrong-AI readings against the first-session readings found no significant difference. [16, 17]
Data as a table
87 optometristsAccuracy, percentPublished 95% interval
Optometrists alone5146 to 56
With AI, no explanation5856 to 60
With AI, feature importance6055 to 64
With AI, scorecard6257 to 66
AI alone, on the 20 study cases80not reported
45 ophthalmologistsPhotographsAccuracy, percent
First reading, unaidedAll 12048.4
Second reading, with AIAll 12059.6
Second reading, with AIPhotographs where the AI was right63.9
Second reading, with AIPhotographs where the AI was wrong47.9

What is available in Australia

The Therapeutic Goods Administration (TGA) publishes a list of AI-enabled devices in the Australian Register of Therapeutic Goods (ARTG). Updated 10 September 2026, it carries eight ophthalmic entries, and the register says what each does [10].

  • TeleMedC's DR Grader screens and grades diabetic retinopathy from colour fundus photographs, in people with diabetes and in people not yet diagnosed.
  • Mediwhale's software reads a fundus image for suspected cataract, glaucoma and retinal disease, and predicts cardiovascular risk.
  • RetInSight's Fluid Monitor and GA Monitor measure fluid volumes and atrophy areas on OCT, to support treatment decisions in neovascular and atrophic age-related macular degeneration.
  • The iCare RETCAD software reads colour fundus photographs for age-related macular degeneration, diabetic retinopathy and glaucoma, as decision support.
  • A Zeiss calculator predicts the post-operative residual refraction for a given biometry and IOL power.
  • Two Eyetelligence products flag referable age-related macular degeneration, diabetic retinopathy, glaucoma and cardiovascular risk from colour fundus photographs. The register notes that the Australian configuration of Assure Plus excludes proliferative diabetic retinopathy and neovascular age-related macular degeneration.

None of the eight is described on the register as deciding without a clinician. Each entry names the health professional it supports. The four FDA-cleared autonomous diabetic retinopathy systems are LumineticsCore, EyeArt, AEYE-DS and iPredict-DR [3, 4]. None is on the TGA list.

The TGA states that its list captures devices where AI or machine learning terms appear "in the product name or intended purpose". It also captures devices whose sponsor reported AI after March 2025. Absence from the list is therefore not absence from the register [10]. SELENA+, a Singapore retinopathy grader, has held ARTG entry 418401 since 17 August 2023 and is not on the list [18]. I therefore searched the public ARTG for each of the four products and its maker on 30 September 2026, and found no matching entry. A name search cannot rule out an entry under a different product name or sponsor [19].

One Australian programme

One documented Australian programme began in Western Australia and now reaches beyond the state. Lions Outback Vision ran Google's ARDA model from a van across nine Pilbara communities between February and August 2024. Seventy-eight patients consented, 10.3 percent of photographs showed referable retinopathy, 8.4 percent were ungradable, and screening rose eleven-fold on 2023 [2].

In October 2025 the partners wrote that "the regulatory steps for implementation of new software as a medical device need to be undertaken" [20]. By April 2026 the spin-out, Ninox Vision, had supplied the tool to a Perth general practice, two clinics in Albany, three Kimberley Aboriginal medical services and a primary care service in Darwin. That list comes from the Australian Medical Association (WA) [21]. Neither source says how the software is regulated today.

The rule that changed in January

The guidance that matters most for a product like mine was rewritten on 29 January 2026. The TGA's guidance on clinical decision support software (CDSS) now states that "an AI-enabled CDSS will not meet the exemption criteria". It reads the third exemption criterion as requiring software to be transparent in how it generates recommendations [22]. Fixed rules do not make a product exempt. All three criteria have to be met. A later issue covers the rules in full.

Conflict of interest, funding and how this was written

I own OptoGuide Pty Ltd, which sells OptoGuide, a clinical decision support product for optometrists. Its clinical recommendations come from fixed rules, and no AI model runs when a clinician uses it. I used AI tools to write its code, to draft its learning cases and to generate its clinical illustrations, which are labelled in the product as AI-generated. It has no ARTG entry. On 30 September 2026 I notified the TGA that it is supplied as exempt clinical decision support software, which is my own determination and not a ruling. It is live in Australia only. Versions for other countries have been researched and are not released. It has no published evaluation, no trial data and no outcome data, so every point in this issue about the gap between validation and outcome applies to it.

No one funded this issue. I have no financial relationship with Eyetelligence, the Centre for Eye Research Australia, Lions Outback Vision, Ninox Vision, Google, or any maker of a diabetic retinopathy system named here. I checked every figure against the paper or the register it is cited to, and the errors are mine. This series will cover the field without promoting OptoGuide, and where a finding bears on rule-based decision support I will say so.

Before I believed OptoGuide changed a patient's outcome, I would want to see referral decisions made with and without it, in clinics I do not run, and followed through to what happened to each patient.

The next issue follows the diabetic retinopathy systems out of their pivotal trials: what held, what moved, who could be photographed and who came back after a positive screen. The issue after that covers the rules for Australian optometrists using AI: the guidance from the Australian Health Practitioner Regulation Agency (Ahpra) and the Code of conduct, the TGA's treatment of software as a medical device, the privacy obligations, including the automated decision-making transparency requirement that starts on 10 December 2026, and where liability sits when a tool is wrong.

A scoping review of this literature is in preparation. Each issue of Second Look is commentary on a subset of the studies. The review is a separate piece of work, and its protocol will be registered before the searches are run.

References

  1. Jan CL, Joseph S, Vingrys AJ, et al. Prospective pragmatic trial of automated retinal photography and AI glaucoma screening in Australian primary care. npj Digit Med. 2025;8(1):386. doi:10.1038/s41746-025-01768-y. PMID 40595445.
  2. Li Q, Drinkwater JJ, Woods K, et al. Implementation of A New, Mobile Diabetic Retinopathy Screening Model Incorporating Artificial Intelligence in Remote Western Australia. Aust J Rural Health. 2025;33(2):e70031. doi:10.1111/ajr.70031. PMID 40110918.
  3. US Food and Drug Administration. Artificial intelligence-enabled medical devices list. Accessed 30 September 2026. https://www.fda.gov/medical-devices/artificial-intelligence-enabled-medical-devices/list-artificial-intelligence-enabled-medical-devices
  4. US Food and Drug Administration. 510(k) K253704, iPredict-DR (iHealthscreen Inc). Decision 2 July 2026. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K253704
  5. Wolf RM, Channa R, Liu TYA, et al. Autonomous artificial intelligence increases screening and follow-up for diabetic retinopathy in youth: the ACCESS randomized control trial. Nat Commun. 2024;15(1):421. doi:10.1038/s41467-023-44676-z. PMID 38212308. NCT05131451.
  6. Abràmoff MD, Whitestone N, Patnaik JL, et al. Autonomous artificial intelligence increases real-world specialist clinic productivity in a cluster-randomized trial. npj Digit Med. 2023;6(1):184. doi:10.1038/s41746-023-00931-7. PMID 37794054. NCT05182580.
  7. Rudnicka AR, Shakespeare R, Chambers R, et al. Automated retinal image analysis systems to triage for grading of diabetic retinopathy: a large-scale, open-label, national screening programme in England. Lancet Digit Health. 2025;7(11):100914. doi:10.1016/j.landig.2025.100914. PMID 41290453.
  8. Wu Y, Qian B, Li T, et al. An eyecare foundation model for clinical assistance: a randomized controlled trial. Nat Med. 2025;31(10):3404-3413. doi:10.1038/s41591-025-03900-7. PMID 40877476. ChiCTR2500095518.
  9. Zhang S, Ran A, Zhou J, et al. An AI-Based OCT System to Detect Diabetic Macular Edema: A Prospective Validation and Noninferiority Randomized Clinical Trial. JAMA. 2026;336(3):215-223. doi:10.1001/jama.2026.7025. PMID 42295755. ChiCTR2300075087.
  10. Therapeutic Goods Administration. AI-enabled medical devices in the ARTG. Published 4 September 2025, updated 10 September 2026. Accessed 30 September 2026 and again 13 October 2026. https://www.tga.gov.au/products/medical-devices/software-and-artificial-intelligence-ai/manufacturing/artificial-intelligence-ai-and-medical-device-software-regulation/ai-enabled-medical-devices-artg
  11. Sharma A, Hussain R, Learoyd AE, et al. Tele-ophthalmology-enabled and artificial intelligence-ready referral pathway for community optometry referrals of retinal disease: HERMES cluster randomised trial with a diagnostic accuracy study. Health Technol Assess. 2025;29(69):1-113. doi:10.3310/QNDF3325. PMID 41424155. ISRCTN18106677.
  12. Jia H, Qian B, Qu Y, et al. AI-based clinician decision support system for diagnosis of inherited retinal diseases: a multicenter, randomized trial. Nat Med. 2026;32(9):3214-3223. doi:10.1038/s41591-026-04545-w. PMID 42498742. NCT06839170.
  13. Yan H, Yi Z, Qilin S, et al. Deep learning-empowered low-cost portable automated refraction system: A solution to the inadequate effective correction rate of refractive errors in resource-limited areas. Ophthalmic Physiol Opt. 2025;45(7):1676-1688. doi:10.1111/opo.70027. PMID 41133975.
  14. Cardon de Lichtbuer P, Declerck W, Rosseel Q, et al. Enhancing patient education in cataract surgery using a conversational artificial intelligence chatbot: pilot randomized controlled trial. J Cataract Refract Surg. 2026;52(5):470-475. doi:10.1097/j.jcrs.0000000000001856. PMID 41408706.
  15. Han R, Acosta JN, Shakeri Z, et al. Randomised controlled trials evaluating artificial intelligence in clinical practice: a scoping review. Lancet Digit Health. 2024;6(5):e367-e373. doi:10.1016/S2589-7500(24)00047-5. PMID 38670745.
  16. Gomez C, Wang R, Breininger K, et al. The explainable AI dilemma under knowledge imbalance in specialist AI for glaucoma referrals in primary care. npj Digit Med. 2025;8(1):706. doi:10.1038/s41746-025-02069-0. PMID 41266539.
  17. Kashiwagi K, Toyoura M, Mao X, et al. Influence of artificial intelligence on ophthalmologists’ judgments in glaucoma. PLoS One. 2025;20(4):e0321368. doi:10.1371/journal.pone.0321368. PMID 40238811.
  18. Therapeutic Goods Administration. ARTG entry 418401: Compliance Management Solutions, EyRIS SELENA+, automated retinopathy analysis system application software. ARTG start date 17 August 2023. Accessed 30 September 2026. https://www.tga.gov.au/resources/artg/418401
  19. Therapeutic Goods Administration. Australian Register of Therapeutic Goods, public search. Searched 30 September 2026 for IDx-DR, LumineticsCore, Digital Diagnostics, EyeArt, Eyenuk, AEYE, iPredict and iHealthscreen. https://www.tga.gov.au/resources/artg
  20. Diabetes WA. How AI is set to transform regional eye care. 23 October 2025. https://www.diabeteswa.com.au/how-ai-is-set-to-transform-regional-eye-care/
  21. Australian Medical Association (WA). GPs offered trial of AI-powered screening tool to battle preventable blindness. 20 April 2026. https://www.amawa.com.au/news/gps-offered-trial-of-ai-powered-screening-tool-to-battle-preventable-blindness/
  22. Therapeutic Goods Administration. Understanding clinical decision support system software regulation. Published 1 February 2021, last updated 29 January 2026. Accessed 30 September 2026. https://www.tga.gov.au/resources/guidance/understanding-clinical-decision-support-system-software-regulation

Second Look is written by Ankit Mathur and published by OptoGuide Pty Ltd. It is not advertising, carries no testimonials and accepts no sponsorship. Figures are drawn from the published values cited beneath each one.

About OptoGuide

OptoGuide is rule-based clinical decision support for therapeutically endorsed optometrists. It is live in Australia and was built by the author of this series.

It runs alongside a consultation, from recognising a presentation through prescribing and referral, and every rule in it is deterministic.